Reference Library-Application Note
K Series Rotors for KII Production Scale Ultracentrifuges
Applications
Responding to production needs for the pharmaceutical industry and now found increasingly in biotechnology, K Series Rotors are used for particle separation via rate zonal separation and isopycnic banding, and can accept corrosive gradients.
Particle types separated include viruses, ribosomes, and other subcellular units with human and animal vaccines, viral vectors, gene therapy, cancer studies, serum and blood fractionation among the important fields of application.
Particles are separated on the basis of their buoyant density. They move under the force of gravity to the band with the same density as their own and remain in this band until recovered by discharging the rotor in liquid fractions after rotation stops.
These rotors can also be used for simple pelleting where the particles are recovered at the end of the run after adhering to the wall of the rotor.
The K3 model completes the linear scale-up range that commenced with the PK3 series of the PKII Pilot Scale UltraCentrifuge. This simplifies protocol development for increasing sample volumes.
The K6 includes a Pre-Clarifier to eliminate the need for a separate step, saving time and increasing convenience and avoiding any possible contamination due to additional sample handling.
Specifications
| Rotor Body – All K Series Rotors |
| |
Max.
RCF
(x g) |
Max.
Speed
(rpm) |
Max.
Radius
(mm) |
Residence Length
(cm) |
Materials |
| |
121,030 |
40,500 |
66 |
76 |
Titanium |
| Rotor Cores |
Materials – Rotor cores can be in Noryl® (most frequent choice) PEEK™ or Titanium |
| |
Type |
Min. RCF
(x g) at
Max. rpm |
Min
Radius (mm) |
Path
Length (mm) |
Volume (Liters) |
 |
K3 |
99,940 |
54.5 |
11.5 |
3.2 |
|
 |
K5 (batch rotor)
|
38,510 |
21 |
45 |
8.4 |
|
 |
K6
.
Built-in
Pre-
Clarifier |
99,940 |
54.5 |
11.5 |
3.2 |
| |
21,090 |
11.5 |
17.5 |
0.35 |
| |
Max. RCF
53,180
|
Max. Rad
29 |
|
|
|
|
 |
K10 |
48,960 |
26.7 |
39.3 |
8.0 |
|
 |
K11 |
120,110 |
65.5 |
0.5 |
0.38 |
|
Linear scale-up
When developing a protocol and starting with a PKII Pilot Scale Centrifuge with PK3-200 rotor, changing to a KII Production Scale UltraCentrifuge gives a linear scale-up factor of 16:1.
Exactly the same time will be used to process 16x as much fluid as for the original volume.
No separation parameter is changed. No protocol development is required.
Massive time savings are achieved with guaranteed results.

|